|
|
|
|
Instruments | EM Receiver | Response Functions
The least count for the MkIII data logger A/D converter is
5 V / 12582911 counts = 3.9736e-07 V / count
The MkIII electric field amplifier has a nominal gain of about 858,000 at
1 s period. The full response curve is shown in the figure below. A table of
the response values can be found at this link: MkIII_Eamp.rsp.
Tests on each amplifier board shows that the calibrated responses vary
by about 1-2%, which is within the specification of the components. This response does
not take into account the standard 10 m dipole length for horizontal
electric field measurements; in other words, the actual electric field measurement
will have another factor of 10 gain from the dipole length.
The BFC-200 induction coil magnetometer with amplifier response is shown
in the figure below. A table of this response can be found at this link: BFC-200_MKIIIamp.rsp.
|
|
|
|
Last updated: Wednesday, 06-May-2009 14:38:31 PDT
email: sconstable@ucsd.edu
|
|